Abstract

An improved test cassette for the integrated Accu-Chek® Mobile system (Roche Diabetes Care GmbH, Mannheim, Germany) has been developed. System accuracy of this improved system was evaluated based on ISO 15197:2013, clause 6.3, for three reagent system lots. According to this standard, at least 95% of the system's measurement results shall be within ±15 mg/dL and ±15% of the results of the comparison method at glucose concentrations <100 mg/dL and ≥100 mg/dL (accuracy criterion A), respectively, and at least 99% of results shall be within consensus error grid zones A and B (accuracy criterion B). In addition, accuracy was evaluated in the hands of users based on ISO 15197:2013, clause 8, with one reagent system lot.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.