Abstract
To evaluate the effect of syringe design and filling technique on the accuracy of anti-vascular endothelial growth factor delivery. Volume output was measured with three syringe designs: a 1.0-mL slip-tip syringe, a 1.0-mL Luer-lock syringe, and a ranibizumab prefilled syringe-using two filling techniques ("upward" and "downward") and two fluids (water and bevacizumab). A total of 300 simulated injections were performed. Accuracy was determined by difference from the intended volume of 50 µ L and by mean absolute percentage error. Volume outputs were significantly different between syringe designs, with mean values of 61.99 ± 4.18 µ L with the 1-mL slip-tip syringe, 57.43 ± 4.95 µ L with the Luer-lock 1-mL syringe, and 51.06 ± 4.74 µ L with the ranibizumab syringe, making the latter the most accurate syringe. There were 37 cases (12.3%) of underdosing below 50 µ L, the majority of which occurred with the ranibizumab syringe. The "downward" technique reduced the occurrence of air bubbles. Intravitreal injections using 1.0-mL syringes are less accurate than using the ranibizumab prefilled syringe, which has a low-volume and low dead-space plunger design. The variability in volume output may result in less predictable treatment response, especially in cases of underdosing, which were more common with the ranibizumab syringe.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.