Abstract

This study aimed to investigate the thermoluminescent properties of ZrO2:Mg irradiated with a 6MV X-ray beam and its potential application in radiotherapy dosimetry. ZrO2 powder was synthesized using the sol-gel method and Mg was used as a dopant. Irradiations were performed with ZrO2:Mg chips located at the center of a 10 × 10cm2 radiation field at a source surface distance of 100cm, below a stack of solid water slabs, at the depth of maximum absorbed dose. The investigated characteristics of the material included linearity with radiation dose, reproducibility, accuracy, sensitivity and fading. Regarding the intrinsic difference of the samples, the glow curves of the investigated ZrO2:Mg chips exposed to 1Gy of 6 MV X-rays exhibited three or four peaks. The ZrO2:Mg samples showed a 47% fading at 24h after irradiation, and the reproducibility of the thermoluminescence reading of ZrO2:Mg for equal irradiation conditions was ± 21%. The thermoluminescence response of the investigated ZrO2:Mg samples to various absorbed doses from 0.5 to 2.5Gy showed a gentle increase of the thermoluminescence intensity with increasing absorbed dose. The obtained results show that ZrO2:Mg is not an appropriate candidate for X-ray photons in radiotherapy, due to low thermoluminescence peak temperature, low reproducibility, low sensitivity to various absorbed doses and significant fading.