Abstract

The Fe3O4/poly(lactide‐co‐glycolide) (PLGA) composites were prepared via a surface grafting technique. Initially, the poly(lactic acid) oligomer was synthesized and surface‐grafted to Fe3O4 nanoparticles. Then, the grafted Fe3O4 particles were compounded with PLGA matrix by a simple solution blending method. The grafted Fe3O4 particles presented enhanced compatibility with PLGA matrix and the composites indicated enhanced dynamic mechanical performance. Electrochemical tests showed that the Fe3O4/PLGA composite coating can help improve the impedance of magnesium samples by 100–300%, and the impedance of the metal may be tunable by altering the components ratio of LA/GA within PLGA matrix. The composites may have potential application for magnesium alloy used in degradable medical implants. POLYM. COMPOS., 37:1369–1374, 2016. © 2014 Society of Plastics Engineers

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.