Abstract

A Pb2+ ion selective membrane electrode based on poly(3,4-ethylenedioxythiophene)-poly(styrenesulfonate) (PEDOT:PSS) Zr(IV) monothiophosphate composite cation exchange material was fabricated using solution casting method. The effect of membrane composition on the proton exchange capacity was investigated by using varying amounts of electroactive material. The membrane with 250 mg of electroactive material and 10 µL of plasticiser exhibited higher proton conductivity. The optimised membrane composition was used for the fabrication of ion selective membrane electrode which exhibited typical Nernstian response towards Pb2+ ions in the concentration range 20.70 gL−1–20.7 µgL−1 (1 × 10–1–1 × 10–7 mol L−1) with a sub-Nernstian slope of 27.429 mV per decade change in Pb2+ ion concentration. The response time of the electrode under study for Pb2+ ions was found to be 11 s and the electrode can be used for 120 days without any considerable divergence in response potential. It can also be successfully used in the pH range from 3.0 to 6.5. It was found selective for Pb2+ ions in the presence of various monovalent, divalent and trivalent interfering metal ions. It was also employed as an indicator electrode in the potentiometric titration of Pb2+ ions using ethylenediaminetetraacetic acid, disodium salt, as a titrant.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call