Abstract
An on‐line system for preconcentration and determination of lead at µg L−1 level is proposed. Amberlite XAD‐2 functionalized with 2‐(2′‐benzothiazolylazo)‐p‐cresol (BTAC‐XAD) packed in a minicolumn was used as sorbent. Chemical and hydrodynamic variables were evaluated and the results demonstrated that lead can be determined using sample pH at 7.0–7.5, elution with 0.050 mol L−1 HCl and a sample flow rate of 6.0 mL min−1. Enrichment factors of 19, 43, and 60 were achieved by using the time‐based technique with preconcentration periods of 60, 120, and 180 s, respectively. Detection limit (3s) was 1.7 µg L−1 when used preconcentration time of 180 s. This procedure was applied to lead determination in natural water samples. Accuracy of procedure was confirmed by lead determination in certified reference materials.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.