Abstract

Synchronization between the left ventricle and a left ventricular assist device (LVAD) may be important for ventricular unloading and coronary perfusion. We assessed the synchrony between cardiac and LVAD cycles by increasing delays in steps of 100 msec throughout the cycle, under conditions of total and partial left ventricular support. We studied 7 healthy minipigs weighing 30-40 kg. A 60-cc Berlin Heart Excor LVAD was implanted and connected to a BCM 1200 console, making it possible to synchronize the LVAD systole and the EKG signal with a prefixed delay. We recorded hemodynamic parameters (including aortic, pulmonary, and left ventricular pressure) and LVAD flow for each delay. Intraventricular pressure during LVAD systole was minimized with delays of around 40-80% of one cycle. In addition, total flow was higher under these conditions. This study shows that the synchronous mode of LVAD operation is feasible. Moreover, a delay in device contraction until the second half of the cardiac cycle optimizes ventricular unloading and may eventually improve myocardial recovery.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.