Abstract
The purpose of this research was to propose synchronous control of the pump flow rate with the natural heart by estimating heartbeat and estimated a pump flow rate without using additional sensors. In this study, we have developed a disturbance observer to estimate heartbeat for estimating the radial thrust, which is the force generating on the impeller by the blood flow in the blood pump. The pump flow rate was also estimated from the radial thrust by experimentally determining the relationship between the pump flow rate and the radial thrust. The validation test was conducted using a simple mock circulation loop, and the heartbeat and the pump flow rate were successfully estimated by the proposed method. Our results concluded that using the heartbeat and pump flow rate estimation system proposed in this study enabled synchronous control with simulated natural heart in a simple mock circulation loop.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Japan Society of Applied Electromagnetics and Mechanics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.