Abstract

Differences in the effectiveness of real-time continuous glucose monitoring (rtCGM) and intermittently scanned continuous glucose monitoring (isCGM) in type 1 diabetes (T1D) are reported. The impact on percent time in range of switching from an isCGM with glucose threshold-based optional alerts only (FreeStyle Libre 2 [FSL2]) to an rtCGM (Dexcom G7) with an urgent low soon predictive alert was assessed, alongside other secondary outcomes including hemoglobin A1c (HbA1c) and other continuous glucose monitoring metrics. Adults with T1D using FSL2 were switched to Dexcom G7 for 12 weeks. HbA1c and continuous glucose data during FSL2 and Dexcom G7 use were compared. Data from 29 participants (aged 44.8 ± 16.5 years, 12 male and 17 female) were analyzed. After switching to rtCGM, participants spent less time in hypoglycemia below 3.9 mmol/L (70 mg/dL) (3.0% [1.0%, 5.0%] vs. 2.0% [1.0%, 3.0%], P = 0.006) and had higher percentage achievement of time below 3.9 mmol/L (70 mg/dL) of <4% (55.2% vs. 82.8%, P = 0.005). Coefficient of variation was lower (39.3 ± 6.6% vs. 37.2 ± 5.6%, P = 0.008). In conclusion, adults with T1D who switched from isCGM to rtCGM may benefit from reduced exposure to hypoglycemia and glycemic variability.