Abstract

In this paper, we present beam pattern tests performed on the SWIPE multi-mode bolometric detector pixel assembly. A 20-mm-aperture horn is coupled to a large detector absorber (10 mm diameter) with a TES sensor located on the side. The pixel assembly has been tested at the bolometer base temperature of 340 mK, inside a custom cryogenic test bed, looking at a Gunn oscillator (128 GHz) located in the far field. We developed a custom cryogenic neoprene absorber, in addition to the stack of standard metal meshes low-pass filters, to reduce the background on the detector at a level similar to the one expected in flight, allowing to measure the main beam of the pixel assembly. The measured FWHM is 21$$^{\circ }$$, slightly narrower than the expected one (24$$^{\circ }$$), due to vignetting produced by the filters stack.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.