Abstract

To figure out the reason for the drawback of the stored waveform inverse Fourier transform (SWIFT) waveform and realize the high-resolution ion isolation on the miniaturized linear ion trap mass spectrometer, we studied the efficiency that ions can be excited under different excitation durations and amplitudes at different frequencies and compared the overlap ratios of the effective excitation frequency bandwidths of the adjacent ions. According to this, we proposed a new coarse-to-fine isolation waveform named SWIFTSIN. By superposing one or more sinusoidal waveforms on the SWIFT waveform and modulating the phases of the superposed sinusoidal waveforms, the generated SWIFTSIN waveform can achieve unit mass isolation on the miniaturized linear ion trap mass spectrometer without reducing the intensity of the target ion. The isolation ability of the SWIFTSIN waveform was verified by isolating a single isotope peak in the mixed samples.