Abstract
Sweeping-window arc therapy (SWAT) is a variation of intensity-modulated radiation therapy (IMRT) with direct aperture optimization (DAO) that is initialized with a leaf sequence of sweeping windows that move back and forth periodically across the target as the gantry rotates. This initial sequence induces modulation in the dose and is assumed to be near enough to a minimum to allow successful optimization, done with simulated annealing, without requiring excessive leaf speeds. Optimal beam weights are calculated analytically, with easy extension to allow for variable beam weights. In this paper SWAT is tested on a phantom model and clinical prostate case. For the phantom, constant and variable beam weights are used. Although further work (in particular, improving the dose model) is required, the results show SWAT to be a feasible approach to generating deliverable dynamic arc treatments that are optimized.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.