Abstract

A novel sweeping under controlled electroosmotic flow scheme was developed for preconcentration and determination of neutral compounds by micellar electrokinetic chromatography (MEKC). An anionic surfactant, sodium dodecyl sulfate (SDS), was added into the buffer for sweeping and separation. By controlled electroosmotic flow (EOF) equal to the counter electrophoretic flow, the surfactants were at an immobile state in capillary. The neutral analytes with sample solution was injected electroosmotically into capillary and swept by SDS micelle for essentially an unlimited volume. The injected sample plug lengths for phlorizin and quercitrin under 18 kV for 70 min were experimentally estimated as 1532 cm, corresponding to 51-fold the effective capillary length. The sweeping under controlled EOF scheme resulted in increased detection factors for phlorizin and quercitrin of 2.3 × 10 4 and 2.1 × 10 4 using 70 min injection relative to a traditional pressure injection. The proposed method has been adopted to analyze trace phlorizin and quercitrin in urine samples successfully.