Abstract

Sutureless end-to-end intestinal anastomoses were successfully constructed in 20 rabbits. A water-soluble intraluminal stent was used to approximate the tissue edges, and the anastomotic seam was lased with 1.06 microns neodymium:YAG (Nd:YAG) laser energy supplied through a hand-held 600-microns gas-cooled noncontact optical fiber. A continuous 5-watt wave of power was applied over periods ranging from 46 to 92 seconds to produce the tissue blanching and shrinkage that indicated a satisfactory tissue weld. Integrity, degree of tissue reaction, and bursting pressures of the anastomoses were compared to those of anastomoses constructed using standard sutured techniques. The results of the two methods were equivalent at 1 to 3 days, but the laser-fused enteric anastomoses showed less inflammatory reaction and greater bursting pressures at 7 and 14 days. Application of a variety of proteinaceous solutions including extracellular matrix materials and epidermal growth factors prior to lasing failed to augment wound healing. We conclude that lasers show promise as reconstructive, in addition to ablative, surgical instruments.