Abstract

To assess the effect of suture material, anchor orientation, and anchor eyelet design on the static loading properties of suture anchors. Biomechanical bench study. Two metallic suture anchors, Mitek GII (Mitek, Westwood, MA) and Corkscrew (Arthrex, Naples, FL) and a bioabsorbable anchor (Biocorkscrew; Arthrex) were tested with single strand of No. 2 Ethibond (Ethicon, Norderstedt, Germany) or No. 2 FiberWire (Arthrex) suture. Suture pull angle was varied through 0 degrees, 45 degrees, and 90 degrees with the anchor rotation angle in either a sagittal or coronal plane. Constructs were tested to failure using an MTS 858 Bionix testing machine (Material Testing Systems, Eden Prairie, MN). Peak loads, stiffness, energy to peak load, and failure modes were determined for all samples. FiberWire showed superior static mechanical properties when compared with single-strand Ethibond over all testing conditions (P < .05). Suture pull angle had a significant effect on load to failure with both metallic anchors but not on the bioabsorbable anchor (P < .05). Suture pull angle and anchor rotation angle play an important role in the failure load of suture when placed in an eyelet. The polyaxial nature of the Biocorkscrew eyelet allows for increased degrees of freedom but introduces failure of the suture eyelet as a new failure mode. The loading direction and placement of the suture anchor plays a role in the performance of the suture anchor-suture complex.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.