Abstract

The purpose of this study was to evaluate survival and clinical outcome of the Scandinavian total ankle replacement (STAR) prosthesis after a minimum of tenyears up to a maximum of 19years. Fifty STAR prostheses in 46 patients with end stage ankle osteoarthritis operated between 1996 and 2006 by the same surgeon (MH) were included. Minimal follow-up was tenyears (median 14.6years, 95% confidence interval [CI] 12.9-16.4). Clinical (Kofoed score) and radiological assessments were taken before the operation and at one, ten (+2), and 16 (±3) years after implantation. The primary endpoint was defined as exchange of the whole prosthesis or conversion to arthrodesis (def. 1), exchange of at least one metallic component (def. 2), or exchange of any component including the inlay (due to breakage or wear) (def. 3). Survival was estimated according to Kaplan-Meier. Further reoperations related to STAR were also recorded. The ten year survival rate was (def. 1) 94% (CI 82-98%), (def. 2) 90% (CI, 77-96%), and (def. 3) 78% (CI 64-87%). The 19-year survival rate was (def. 1) 91% (CI 78-97%), (def. 2) 75% (CI 53-88%), and (def. 3) 55% (CI 34-71%). Considering any re-operations related to STAR, 52% (26/50) of prostheses were affected by re-operations. Mean pre-operative Kofoed score was 49, which improved to 84 after one year (n=50), to 90 after ten years (n=46), and to 89 after 16years (n=28). The survival rate for def. 1 and 2 was high. However, re-operations occurred in 52% of all STAR prosthesis. Retrospective cohort study, evidence Level 4.