Abstract

Surgical smoke evacuators may reduce the concentration of alcohol vapors from skin preparations at the site of electrosurgical device activation, decreasing operating room fire risk. Our aim was to compare the incidence of flames with and without smoke evacuation in a porcine ex vivo model. A monopolar device was activated after application of either 70% isopropyl alcohol/2% chlorhexidine gluconate (CHG-IPA) or 74% isopropyl alcohol/0.7% iodine povacrylex (iodine-IPA) skin preparations. No suction was compared with standard wall suction and 2 monopolar devices with built-in smoke evacuators. Flames were generated in 60% of CHG-IPA and 47% of iodine-IPA tests without suction. Wall suction did not significantly reduce fires (CHG-IPA 43% vs. 60%, P=0.30; iodine-IPA 57% vs. 47%, P=0.61). Use of both smoke evacuation devices reduced fires for CHG-IPA (17% vs. 60%, P=0.001 and 20% vs. 60%, P=0.004) but not for iodine-IPA. Smoke evacuation devices reduce fire risk when used with a chlorhexidine-alcohol skin preparation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.