Abstract

To investigate the safety and efficacy of a newly-developed wide-field dual-array suprachoroidal-transretinal stimulation (STS) prosthesis in middle-sized animals. The prosthesis consisted of two arrays with 50 to 74 electrodes. To test the feasibility of implanting the prosthesis and its efficacy, the prosthesis was implanted for 14days into two rabbits. Optical coherence tomography (OCT) and ophthalmoscopy were performed 7 and 14days after the implantation. Then the rabbits were euthanized, eyes were enucleated, and the posterior segment of the eye was examined histologically. In a second experiment, the arrays were implanted into two cats, and their ability to elicit neural responses was determined by electrically evoked potentials (EEPs) at the chiasm and by optical imaging of the retina. All arrays were successfully implanted, and no major complications occurred during the surgery or during the 2-week postoperative period. Neither OCT nor ophthalmoscopy showed any major complications or instability of the arrays. Histological evaluations showed only mild cellular infiltration and overall good retinal preservation. Stimulation of the retina by the arrays evoked EEPs recorded from the chiasm. Retinal imaging showed that the electrical pulses from the arrays altered the retinal images indicating an activation of retinal neurons. The thresholds were as low as 100μA for a chiasm response and 300μA for the retinal imaging. Implantation of a newly-developed dual-array STS prosthesis for 2weeks in rabbits was feasible surgically, and safe. The results of retinal imaging showed that the dual-array system was able to activate retinal neurons. We conclude that the dual-array design can be implanted without complication and is able to activate retinal neurons and optic nerve axons.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call