Abstract
A 70-year-old man with dilated cardiomyopathy underwent left ventricular assist device (LVAD) implantation, using a HeartWare ventricular assist device, as a bridge to candidacy. After 26 months, computed tomography (CT) angiography indicated stenosis in the LVAD outflow graft; however, the patient was asymptomatic, prompting a decision to manage his condition with close monitoring. Ten months later, the patient presented with dizziness and low-flow alerts. Subsequent CT angiography revealed a critical obstruction involving the entire LVAD outflow graft. The patient underwent emergency surgery, during which an organized seroma causing the graft obstruction was found between a wrapped expanded polytetrafluoroethylene (ePTFE) graft and a Dacron outflow graft. The covering of the outflow graft was removed, along with the organized seroma. Following removal of the ePTFE wrap and decompression of the outflow graft, normal LVAD flow was reestablished. The practice of wrapping the outflow graft with synthetic material, commonly done to facilitate later redo sternotomy, may pose a risk for outflow graft obstruction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.