Abstract

Sterilization of the NiTi alloy in boiling water or steam causes passivation, which results in an amorphous 3.5 nm thick TiO2 layer on the surface. Between the surface and the matrix a transition layer of Ni2O3 and NiO was observed, using the X-ray photoelectron spectroscopy. Differences in sterilization conditions affect the amount of metallic nickel on the surface.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.