Abstract

To evaluate and compare tapping mode and contact mode AFM measurements of surface roughness, images of quartz and mica were acquired by both methods and the height distributions and variance correlation functions analyzed. Significant deviation from the expected Gaussian profiles for the height distributions were observed for contact mode images of quartz but not for tapping mode images. Additionally, variance correlation functions were found to be highly scan size dependent for contact mode images and scan size invariant for tapping mode images. One possible explanation for the observed differences is that the scan speed limit is exceeded in contact mode for linear scan velocities as low as 0.5 μm/s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.