Abstract

Purpose: To determine the surface characteristics of porcine corneal lenticules after Femtosecond Lenticule Extraction. Methods: The Carl Zeiss Meditec AG VisuMax® femtosecond laser system was used to create refractive corneal lenticules on 10 freshly isolated porcine eyes. The surface regularity on the corneal lenticules recovered was evaluated by assessing scanning electron microscopy images using an established scoring system. Results: All specimens yielded comparable score results of 5–7 points (SD = 0.59) per lenticule (score range minimum 4 to maximum 11 points). Surface irregularities were caused by tissue bridges, cavitation bubbles or scratches. Conclusion: The Femtosecond Lenticule Extraction procedure is capable of creating corneal lenticules of predictable surface quality. However, future studies should focus on the optimization of laser parameters as well as surgical technique to improve the regularity of the corneal stromal bed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.