Abstract

In this paper, the biocompatibility of a medical-grade stainless steel coated with sol-gel derived, nanostructured inorganic ZrTiO4 and hybrid ZrTiO4-PMMA thin films is correlated with surface characteristics. The surfaces of the samples are characterized by atomic force microscopy, the sessile drop technique, and electrochemical corrosion experiments. The viability of adult human mesenchymal stem cells on the surfaces after one day of culture is also assessed quantitatively and morphologically. According to the results, both of the coatings improve the hydrophilicity, corrosion resistance, and thereby cytocompatibility of the substrate. Despite the higher corrosion protection by the hybrid coating, the sample coated with the inorganic thin film exhibits a better cell response, suggesting the domination of wettability. In summary, the ZrTiO4-based sol-gel films can be considered to improve the biocompatibility of metallic implants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.