In this study, we have successfully demonstrated a nanostructured phenylboronic acid-grafted poly(3,4-ethylenedioxythiophene), poly(EDOT-PBA), platform for fast and sensitive glucose monitoring. The poly(EDOT-PBA) films of well-organized tubular nanostructures can be fabricated by direct electropolymerization without templates. Compared to the smooth poly(EDOT-PBA), the nanotubular poly(EDOT-PBA) shows enhanced glucose sensitivity and a different adsorption process of bovine serum albumin (BSA). Besides, the BSA blocking and low concentration of fructose and galactose do not affect the sensitivity of this platform. Both quartz crystal microbalance (QCM) and electrochemical impedance spectroscopy (EIS) methods are used and compared for glucose monitoring by applying nanotubular poly(EDOT-PBA) as conductive substrates. Compared to QCM analysis, EIS has a higher sensitivity to glucose and the detection limit is about 50 μM. Besides, the binding with glucose on poly(EDOT-PBA) is highly reversibly. On the basi...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.