Abstract

Photobleaching can lead to significant errors in frequency-domain fluorescence lifetime imaging microscopy (FLIM). Existing correction methods for photobleaching require additional recordings and processing time and can result in additional noise. A method is introduced that suppresses the effects of photobleaching without the need for extra recordings or processing. Existing bleach correction methods and the method introduced in this report whereby the recording order of the phases is permuted were compared using numerical simulations. Certain orders were found to make measurements virtually insensitive to photobleaching. At 12 recordings, errors in measured phase and modulation depth decreased by a factor 512 and 393, respectively, compared to recordings using sequential recording order. The optimal order is independent of modulation depth, phase, and extent of photobleaching. Thus, the same order can be used for practically all situations. Application of the method in FLIM measurements of EYFP-transfected HeLa cells was found effectively to suppress photobleaching induced artifacts. In view of the ease of implementation, its inherent robustness, and the possibility to still apply existing correction methods afterward, there is no good reason not to use the permuted recording order presented in this report instead of a sequential order.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.