Abstract

Magnetopneumography is a noninvasive method for examining contamination of metal workers' lungs by ferromagnetic dusts. Our aim is to estimate not only total amount but also spatial distribution of the dust. We use fluxgate magnetometer for measuring remanent field of contaminated volume (current experiments are mostly on phantoms). For earlier experiments we used a simple 6-channel magnetometer with improvised probe holder and we used multiple first order gradients. However, geometrical and other instabilities forced us to upgrade both the mechanical assembly and the fluxgate electronics. The environmental noise from relatively close sources is apparent even in first order gradients and thus - now with more stable electronics - we evaluated second order gradients in order to obtain a better signal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.