Abstract

Alternate ascending/descending directional navigation (ALADDIN) is a new imaging technique that provides interslice perfusion-weighted and magnetization transfer (MT) asymmetry images. In this article, we investigated the effects of gradient imperfections on ALADDIN MT asymmetry (MTA) signals. Subtraction artifacts increasing with readout offsets were detectable in ALADDIN MTA images from an agarose phantom but not from a water phantom. Slice-select offsets had no significant effect on the artifacts in MTA. The artifacts were suppressed by averaging signals over the readout gradient polarities independent of scan parameters. All these results suggested that the subtraction artifacts were induced by readout eddy currents. With suppression of the artifacts, ALADDIN signals in human brain and skeletal muscle varied less with scan conditions. Percent signal changes of MTA in human skeletal muscle (0.51 ± 0.11%, N = 3) were about 30% of those in white matter. The new averaging scheme will allow for more accurate MTA imaging with ALADDIN, especially at off-center positions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.