Abstract

In combination with a single-crystal diamond anvil cell (DAC), a polycapillary half-lens (PHL) re-focusing optics has been used to perform high-pressure extended X-ray absorption fine-structure measurements. It is found that a large divergent X-ray beam induced by the PHL leads the Bragg glitches from single-crystal diamond to be broadened significantly and the intensity of the glitches to be reduced strongly so that most of the DAC glitches are efficiently suppressed. The remaining glitches can be easily removed by rotating the DAC by a few degrees with respect to the X-ray beam. Accurate X-ray absorption fine-structure (XAFS) spectra of polycrystalline Ge powder with a glitch-free energy range from -200 to 800 eV relative to the Ge absorption edge are obtained using this method at high pressures up to 23.7 GPa, demonstrating the capability of PHL optics in eliminating the DAC glitches for high-pressure XAFS experiments. This approach brings new possibilities to perform XAFS measurements using a DAC up to ultrahigh pressures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.