Abstract

This report gives the results of a comparison of pressure standards of three National Metrology institutes in the range of 0.7 MPa to 7 MPa. This comparison was piloted by UME and was carried out from February 2017 to March 2019. This work is a part of the GULFMET project and is registered as a supplementary comparison GULFMET.M.P-S1. The transfer standard used was a digital pressure gauge, model 745-1K, serial number 114944, manufactured by Paroscientific Inc, with a resolution of 1 hPa. The reference values have been determined from the weighted mean of the deviations which is a measurement error of the test instrument reported by the participants for each specified pressure.Main textTo reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.