Abstract
A capillary scale reverse phase liquid chromatography (LC) system using a super hot water eluent is described. The system, constructed in-house from readily available components, has been shown to operate at temperatures as high as 370 °C and pressures in excess of 10 000 psi. The capability of the system is demonstrated with the separation of a mixture of polar and non-polar benzene derivatives on polybutadiene and elemental carbon modified zirconia packings with or without temperature gradients. Six benzene derivatives can be separated in ∼2 min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.