Abstract

A superfluid helium insert was developed for cryogenic microscopy of millimeter-sized specimens. An optical-interferometric position sensor, cryogenic objective mirror, and piezo-driven cryogenic stage were fixed to an insert holder that was immersed in superfluid helium. The single-component design stabilized the three-dimensional position of the sample, with root-mean-square deviations of (x, lateral) 0.33nm, (y, lateral) 0.29nm, and (z, axial) 0.25nm. Because of the millimeter working range of the optical sensor, the working range of the sample under the active stabilization was (x, y) 5mm and (z) 3mm in superfluid helium at 1.8K. The insert was used to obtain the millimeter-sized fluorescence image of cell nuclei at 1.8K without a sample exchange.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call