Abstract
Atmospheric simulation chambers play vital roles in the validation of chemical mechanisms and act as a bridge between field measurements and modeling. Chambers operating at atmospheric levels of OH radicals (106-107 molecule/cm3) can significantly enhance the possibility for investigating the discrepancies between the observation and model predications. However, few chambers can directly detect chamber OH radicals at ambient levels. In this paper, we report on the first combination of a superconducting magnet with midinfrared Faraday rotation spectroscopy (FRS) for real time in situ measurement of the OH concentration in an atmospheric simulation chamber. With the use of a multipass enhanced FRS, a detection limit of 3.2 × 106 OH/cm3 (2σ, 4 s) was achieved with an absorption path length of 108 m. The developed FRS system provided a unique, self-calibrated analytical instrument for in situ direct measurement of chamber OH concentration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.