Abstract
The need for mapping skin doses during fluoroscopically-guided X-ray procedures has been described by a number of institutions and experts. Different large photographic or X-ray films placed on the patient's skin have been found to be useful for recording doses up to 1.0-2.0 Gy - depending on the film - and up to 15 Gy using radiochromic films. Though the upper limit of the film sensitivity is seldom exceeded during interventional procedures, the main disadvantage of the X-ray films is still the excessive sensitivity for long, high dose procedures. Radiochromic films show poor definition for doses below 0.5 Gy and are expensive. The goal of the present paper is to analyse the possibilities of using common resin-coated photographic paper for this purpose. Sensitometric curves obtained with different paper types processed in conventional X-ray film automatic processors demonstrate that some of them can be used with better results than X-ray films at a very low cost. Doses from about 10 mGy to near 3.0 Gy can be measured with good accuracy using a variety of glossy photographic papers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.