Abstract

The aim of this study was to inspect the usefulness of relative artifact index (AIr), which divided artifact index (AI) by standard deviation of the noise image for noise-independent evaluation of the streak artifact in computed tomography images. A water phantom without/with a cylindrical phantom filled with diluted contrast medium was scanned with different tube voltages (100/120/140 kV) and radiation doses (5/10/20 mGy), then images were reconstructed with different kernels (B10/30/50f). AI, location parameter in Gumbel method and AIr were measured in each condition and compared. The higher tube voltage or radiation dose or lower spatial resolution kernel, the lower quantitative values were presented by both AI and Gumbel method. AIr showed quantitative values independent of radiation dose and kernel, and substantial artifact amounts affected only by tube voltage. Our results showed AIr can evaluate quantitative artifact amount independent of image noise.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.