Abstract

Purpose: Measure scatter and leakage dose in critical, extracranial organs for patients undergoing Leksell Gamma Knife Radiosurgery.Method and Materials: A stereotactic frame was attached to the head of an anthropomorphic Rando phantom. Using a fiducial box to determine the coordinates of the target (center of the Rando head), a CT scan was taken and registered in the GammaPlan treatment planning system where a dose prescription of 25Gy to the 50% isodose line was applied to the target site. The plan was generated for the 18 mm collimator size helmet with a single shot run using the automatic positioning system (APS). Multiple (2 to 5) 2″×2″ Gafchromic EBT films were placed between each slice of the Rando body phantom, from the neck to pelvic region (phantom slice 8 to 31), and analysis was done using a Vidar VXR‐16 scanner with the RIT113 Version5.1 analysis software. An HD dose will also decrease with increasing distance from the focal point. The doses are low but may be considered for younger patients with longer life expectancy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.