Abstract

Purpose: AcurosXB and AAA dose calculation algorithms are assessed for accuracy in RapidArc stereotactic (small field) treatments. Methods: AcurosXB (engineering pre‐release 11.0.03) and AAA (release 10.0.28) as implemented in Varian Eclipse planning system were configured and analyzed. To assess the baseline accuracy, small open fields were considered first. Spot size setting was varied for to 0.5, 1, and 2mm. Calculation grid was 1mm. Field settings were 3×3, 2×2, 1×1 and 0.8×0.8 cm2, 6MV beam from a Clinac2100iX. Profiles, PDD, and output factors were measured with PTW diamond detector (detector size: 4mm2, thickness 0.4mm) and compared to calculations. Additionally, four RapidArc cases were optimized, calculated and delivered with jaw settings 3×3, 2×2 and 1×1 cm2, the last having two plans, one with high (H) and one with low (L) modulation. Each plan contained one partial arc (gantry 110° to 250°), and collimator 45°. Dose to isocenter was measured in a PTW Octavius phantom and compared to calculations. Results: Open square fields: penumbrae from open field profiles were in very good agreement with diamond for 1 mm spot size setting for Acuros, and between 0.5 and 1mm for AAA. Maximum MU difference between calculations and measurements was 1.7% for Acuros XB (0.2% for fields greater than 1×1cm2) with 0.5 or 1mm spot size. Agreement for AAA was within 0.7%, 2.8% for 0.5, 1mm spot size, respectively. RapidArc cases: computed doses evaluated in an isocentric 6mm diameter sphere differ from measurements of −0.1, −1.1, 4.1, −1.0% for AcurosXB calculations and 1mm spot size, and of −0.6, −0.5, 1.8, −2.7% for AAA, respectively for 3×3, 2×2, 1×1H, 1×1L Rapidarc plans. Conclusions: RapidArc dose caclulations with AAA or AcurosXB algorithms can be considered a safe solution for stereotactic treatments. The present work was partially supported by a Grant from Varian Medical Systems, Palo Alto, CA, Dr. L. Cozzi acts as Scientific Advisor to Varian Medical Systems and is Head of Research and Technological Development to IOSI, Bellinzona.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.