Abstract
The paper investigates a possibility of pMOS dosimeter re-use for the measurement of gamma-ray irradiation. The dosimeters were irradiated to the dose of 35 Gy, annealed at room and elevated temperatures, after which they were irradiated again to the same dose value. Changes in the threshold voltage shift during those processes were followed, and it was shown that their re-use depends on a gate polarization during irradiation. For the gate polarization of 5 V during irradiation the pMOS dosimeters can be re-used for measurements of the irradiation dose after annealing without prior calibration. The pMOS dosimeters with the gate polarization during irradiation of 2.5 V can also be re-used for irradiation dose measurements but they require calibration. It is shown that for their re-use it is necessary to anneal the pMOS dosimeter so that the fading is higher than 50%.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.