Abstract

Conducting a general anaesthetic within a magnetic resonance imaging (MRI) suite poses many problems for the anaesthetist. Ferromagnetic substances are contained in most anaesthetic and monitoring equipment. Their presence within the magnetic field may cause hazards and artifacts during imaging. This report describes the testing and use of a free-standing Fortec II vaporizer within an MRI suite. The Fortec II vaporizer's function was altered depending upon its distance from and orientation to the magnetic field. The MRI images were not affected by this vaporizer's presence within the MRI suite. We conclude that an inhalational anaesthetic can be administered, using a pretested free-standing anaesthetic vaporizer and a Bain circuit, within the magnetic field of a magnetic resonance imager.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.