Abstract

Stent migration from the delivery balloon catheter is a rare but serious complication during percutaneous coronary intervention, particularly when a part of the stent stretches into the aorta. We report an unusual case of stent migration treated with a combination of a gooseneck snare and rotablation. A part of the stent was overstretched and unrolled into the aorta and the rest of the stent remained implanted in the coronary artery. The stent was captured with a gooseneck snare but could not be retrieved because it was connected to a stent remnant implanted in the coronary artery. The stent strut was cut with rotablation, and the stent was successfully removed through the femoral sheath.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.