Abstract

We report the successful retrieval of an impacted mechanical lithotripsy basket. In a patient with two large common bile duct stones, the basket with the entrapped stone was impacted within the mid-common bile duct. We then attempted to use another mechanical lithotripter; however the central wire of the basket fractured at the handle portion. Grasping a few wires of the impacted basket with rat-tooth forceps allowed the wires of the basket to slip away from the stone. The present report describes the safe and effective use of rat-tooth forceps in the management of an impacted lithotripter basket.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.