Abstract

Duodenal stenting has been gradually established in recent years because it is less invasive than standard surgical procedures and produces a rapid therapeutic response. For palliation of both duodenal and biliary stenoses, double stenting may be performed. Duodenal stents offer a great advantage in allowing endoscopic retrograde cholangiopancreatography (ERCP) without the need for balloon dilation. When biliary stent dysfunction occurs, the patient undergoes diagnostic and/or therapeutic ERCP across the duodenal stent. We encountered a duodenal stent fracture in a patient who required repeated ERCPs for stent dysfunction. Duodenal stent fractures have not previously been reported. The damaged stent was successfully repaired by using a cutting wire filament and placing another duodenal stent coaxially with the first. Clinicians should be aware of the possibility of stent fracture following endoscopic procedures, such as an ERCP, that require passage through the stent. The procedure described in this report would be of significant benefit if a gastrointestinal stent is fractured and occluded by a broken part.