Abstract
Nitinol stents are thought to exhibit reduced occurrence of artifacts and may be suitable for magnetic resonance imaging (MRI) evaluation of stent localization and in-stent patency even in coronary-sized stent grafts. A 54-year-old male patient presented with a large coronary post-stenotic aneurysm of the right coronary artery (RCA) beside significant stenoses of the left circumflex coronary artery (LCX) and the left anterior descending coronary artery (LAD) with aneurysm formation. After implantation of stent grafts to the LAD and LCX, two polymermembrane-covered nitinol stent grafts were placed into the RCA. A control MR examination 7 days following the RCA intervention showed successful occlusion of the former aneurysm, no post-interventional endoleak, and bright signal within the stent indicating stent patency. Thus, coronary MRI after nitinol stent implantation in coronary aneurysms is feasible for post-interventional early imaging control at least as far as the exclusion of possible endoleaks is concerned.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The International Journal of Cardiovascular Imaging
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.