Abstract
ObjectivesAlthough rare, incidents of broken/dislodged fragment of various angioplasty hardwares, including catheters, guidewires, angioplasty balloons, and stents, are being increasingly reported in recent years. Since these broken fragments may cause life-threatening consequences for a patient, it is vital for an interventional cardiologist to be acquainted with different retrieval techniques. Here, we are reporting our observations of several incidents of device dislodgement/fracture during cardiac interventions and their retrieval using simple balloon method.MethodsWe present a study of eight patients in whom we attempted to retrieve dislodged/fractured cath-lab hardwares during cardiac interventions, using simple balloon method. These cases include two cases of balloon, three cases of stent, and three cases of guidewire dislodgement/fracture.ResultsFractured/dislodged cath-lab hardwares were successfully retrieved using a simple balloon method in six out of eight cases (75%), without any need of other retrieval hardwares. We observed no major complications in any patient.ConclusionsThe balloon-assisted retrieval method is a simple, safe, and cost-effective way to avoid complications of endothelial injury, myocardial infarction, emergency coronary artery bypass graft, and sudden cardiac death. This study, particularly the context of retrieval technique used in each case, will offer valuable information to fellow interventional cardiologists.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.