Abstract
From February 1988 until August 1991, 28 patients with prior coronary artery bypass grafting (CABG) received implantable cardioverter defibrillator (ICD) therapy via a subxiphoid approach. Only one patient required conversion to a median sternotomy incision. The mean defibrillation threshold was 11.9 +/- 4.4 J. The mean R wave was 8.2 +/- 3.7 mV. One perioperative death occurred due to heart failure (mortality rate 1/28 [3.50%]). No patient required reexploration for bleeding. The subxiphoid method for ICD electrode implantation is safe and reliable in patients with prior CABG surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.