Abstract
It remains uncertain whether prosthetic ring size should be used interchangeably with measured left ventricular outflow tract (LVOT) diameter in the continuity equation to estimate the aortic prosthetic valve area by transthoracic Doppler echocardiography. To determine the difference in area caused by this substitution, the area of the prosthetic valve was calculated in 143 patients with aortic bioprostheses by use of the standard continuity equation with the measured LVOT diameter (LVOT method) and then with the bioprosthetic size (size method). Compared with known in vitro prosthetic valve areas, the LVOT method (r = 0.86; standard error of the estimate +/- 0.16 cm2; p < 0.001) was more accurate than the size method (r = 0.74; standard error of the estimate +/- 0.40 cm2; p < 0.001). The prosthetic valve area estimated by the size method overestimated the area estimated by the LVOT method by an average of 15% +/- 23% (p < 0.001). This difference in area between the two methods has increased with the interval since implantation of the bioprosthesis (p = 0.01). It is concluded that prosthetic size should not be used instead of LVOT diameter during calculation of aortic prosthetic valve area. This restriction is particularly important in patients with older bioprosthesis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the American Society of Echocardiography
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.