We describe a sensitive and specific method for the measurement of buprenorphine in human plasma. The method involves a structural analog as an internal calibrator, careful control of pH during sample extraction to maximize drug recovery, and back-extraction into acid followed by reextraction to eliminate endogenous interferences. After evaporation, sample residues are derivatized with heptafluorobutyric anhydride and analyzed by separation on a fused-silica polymethylsiloxane capillary column and electron-capture detection. Calibration curves were linear in the ranges 0.1-2.0 micrograms/L and 2.0-20 micrograms/L, with within-run CVs of 9.7% at 0.1 microgram/L to 5.0% at 20 micrograms/L, and total CVs of 15.9% at 0.1 microgram/L to 6.5% at 10 micrograms/L. The limit of quantification was 0.1 microgram/L. The method was utilized in studies to determine the absolute bioavailability of sublingual doses of 2 mg of buprenorphine in 1 mL of 300 mL/L ethanol and the bioequivalence of sublingual 8-mg tablet and 300 mL/L ethanol solution formulations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.