Abstract

The purpose of this study is to identify regions of spinal column in which more than three contiguous vertebrae can be reliably and quickly aligned within 1 mm using a 6‐degree (6D) couch and full body immobilization device. We analyzed 45 cases treated over a 3‐month period. Each case was aligned using ExacTrac x‐ray positioning system with integrated 6D couch to be within 1° and 1 mm in all six dimensions. Cone‐Beam computed tomography (CBCT) with at least 17.5 cm field of view (FOV) in the superior–inferior direction was taken immediately after ExacTrac positioning. It was used to examine the residual error of five to nine contiguous vertebrae visible in the FOV. The residual error of each vertebra was determined by expanding/contracting the vertebrae contour with a margin in millimeter integrals on the planning CT such that the new contours would enclose the corresponding vertebrae contour on CBCT. Submillimeter initial setup accuracy was consistently achieved in 98% (40/41) cases for a span of five or more vertebrae starting from T2 vertebra and extending caudally to S5. The curvature of spinal column along the cervical region and cervicothoracic junction was not easily reproducible between treatment and simulation. Fifty‐seven percent (8/14) of cases in this region had residual setup error of more than 1 mm in nearby vertebrae after alignment using 6D couch with image guidance. In conclusion, 6D couch integrated with image guidance is convenient and accurately corrects small rotational shifts. Consequently, more than three contiguous vertebrae can be aligned within 1 mm with immobilization that reliably reproduces the curvature of the thoracic and lumbar spinal column. Ability of accurate setup is becoming less a concern in limiting the use of stereotactic radiosurgery or stereotactic body radiation therapy to treat multilevel spinal target.

Highlights

  • Spinal stereotactic radiosurgery (SRS) or stereotactic body radiation therapy (SBRT) delivers high dose in one or a few fractions is increasingly used to maximize local tumor control and pain relief.[1]

  • The Cone-Beam computed tomography (CBCT) had a range of at least 17.5 cm in the superior–inferior direction in the study. This scan covered five to nine contiguous vertebrae depending on the patient size, disease location, and spinal column curvature

  • Four cases showed initial setup accuracy can be achieved within 1 mm for a span of five or more vertebrae centered on T2 vertebra

Read more

Summary

Introduction

Spinal stereotactic radiosurgery (SRS) or stereotactic body radiation therapy (SBRT) delivers high dose in one or a few fractions is increasingly used to maximize local tumor control and pain relief.[1]. In our institution’s current practice, it has been mostly limited to three or less contiguous vertebrae. Such a limitation is due to the caution and uncertainties of normal tissue tolerance in spinal SRS/SBRT and great concern on the technique ability to align long/ large target accurately and reliably. Even a small setup error of 2 mm or 2° can significantly affect both the target and spinal cord dose.[2] The effect of small rotational error is more significant when treating large target spanning three or more vertebral bodies, since such error could cause large displacement at distance from the isocenter. Rotational error is especially challenging to correct with a conventional treatment couch

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.