Abstract
Advice to patients following grommet insertion and waterproofing can vary from clinician to clinician. A laboratory based experiment was performed to determine at what depth water contamination would occur through various grommet tubes. A novel experimental ear model was developed using an artificial tympanic membrane and ventilation tubes. Water contamination was identified using an effervescent solid that reacts when in contact with water. Measures of dispersion were used to describe the results. The average depth of water contamination was: 19.64 mm (range = 11-33 mm, standard deviation = 5.55 mm) using a Shepard grommet; 20.84 mm (range = 18-26 mm, standard deviation = 1.97 mm) with a titanium grommet; and 21.36 mm (range = 18-33 mm, standard deviation = 3.03 mm) using a T-tube. Water contamination was possible at depths of 11-33 mm. The average pressure at water effervescent activation was 0.20 kPa. Submersion underwater at any depth with grommets is likely to lead to middle-ear contamination. These findings are concordant with clinical studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.