Abstract
To introduce a novel methodology for subjective refraction based on power vectors with a conventional phoropter. A conventional phoropter was used to measure power vector components of refraction (M, J0 and J45) directly by using the sphere power (for M measurement) and the cylinder power combined with the Jackson cross-cylinders (for J0 and J45 measurements). Conventional subjective refraction was also performed, and this result was mathematically transformed into power vector notation for comparison purposes. Visual acuities with the conventional prescription and the quasi-vector-based prescription were compared. Refractive error from 40 healthy participants was measured by conventional and quasi-vector-based subjective refraction. No differences were found between methods for any of the power vector components of refraction (p > 0.21 in all cases). The visual acuity achieved with the prescriptions yielded similar values (p = 0.85). Subjective refraction can be measured directly in power vector notation using a conventional phoropter without any additional adaptation and computation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.