Abstract

Postmortem examination two weeks following implantation of a demand pacemaker with sutureless electrodes demonstrated that (1) the loops of intrapericardial electrode leads indented the myocardium and produced a circumscribed area of subepicardial necrosis, and (2) one of the electrode coils protruded into the right ventricular cavity. The use of extrapericardial electrode leads and left ventricular implantation sites should avoid these problems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.